← Database

CVE-2024-9832

CRITICAL CVSS: 9.3 Published: 2024-11-14 Modified: 2026-06-17

Description

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

Affected Platforms

Unknown

Weakness Type

CWE-307

References